Entry
| Date
| Title
| Site
| Author
| #Graphics
|
15 | Sun 02-Feb-1992 | replace 10M psych. Exchanged unit 58 fo | none | AA
|
19 | Mon 03-Feb-1992 | 10m RH still odd | none | twh
|
21 | Mon 03-Feb-1992 | PSYC CHANGE, 10m with 2m: | none | srs
|
22 | Mon 03-Feb-1992 | Moved pressure port. | none | AA
|
64 | Sun 09-Feb-1992 | EGG SIGNALS: | none | srs
|
69 | Mon 10-Feb-1992 | 2m HUMITTER | none | aa
|
71 | Mon 10-Feb-1992 | EGG t. | none | aa
|
88 | Wed 12-Feb-1992 | EGG | none | aa
|
106 | Tue 18-Feb-1992 | Removed 10m psychrometer and cleaned hu | none | gdm
|
113 | Wed 19-Feb-1992 | Replaced psyc.1am with PAM psyc | none | gdm
|
119 | Fri 21-Feb-1992 | pam IC intercompare started | none | clm
|
128 | Mon 24-Feb-1992 | Cleaned and replaced filter on 2m humit | none | gdm
|
131 | Mon 24-Feb-1992 | Removing the Asman intercomparison unit | none | acd
|
133 | Tue 25-Feb-1992 | Attached PAM psychrometer #35 to psyc.1 | none | gdm
|
137 | Wed 26-Feb-1992 | Maintainence of 10m humitter (unit 51) | none | gdm
|
142 | Thu 27-Feb-1992 | SWITCHING HUNCHROMETERS 1am for 10m | none | acd
|
146 | Fri 28-Feb-1992 | Replaced PAM pysc #51 with #55 | none | gdm
|
149 | Sun 01-Mar-1992 | IC unit 55: | none | srs
|
161 | Tue 03-Mar-1992 | IC unit 55: | none | srs
|
162 | Tue 03-Mar-1992 | INC unit 55: | none | srs
|
174 | Fri 06-Mar-1992 | 7M PSYC DOWN & 2M RH ?: | none | srs
|
180 | Sun 08-Mar-1992 | PSYC IC UNIT: | none | srs
|
183 | Sun 08-Mar-1992 | 2M PSYC: | none | srs
|
191 | Mon 09-Mar-1992 | 2M RH MISBEHAVING: | none | srs
|
192 | Mon 09-Mar-1992 | 2M PSYC FIX: | none | srs
|
198 | Tue 10-Mar-1992 | 2M PSYC STATUS: | none | srs
|
202 | Wed 11-Mar-1992 | 2M RH RUNNING: | none | srs
|
208 | Sat 14-Mar-1992 | PSYC IC UNIT REMOVED: | none | srs
|
214 | Sun 15-Mar-1992 | PSYC IC TEST: | none | srs
|
233 | Wed 21-Jul-1993 | Don't trust some temperature/humidity p | none | spo
|
- 15: PSYC , Site none, Sun 02-Feb-1992 19:56:48 GMT, replace 10M psych. Exchanged unit 58 fo
replace 10M psych. Exchanged unit 58 for unit 35.
bad humidity readings, too high.
- 19: PSYC , Site none, Mon 03-Feb-1992 15:40:57 GMT, 10m RH still odd
10m RH still odd
We note that the 10m RH still appears to still be acting
as before, that is it still reads higher than would be
expected from a normal profile, particularly during the
day. While this could perhaps be real, we are going to
switch the 10m with a lower level to further check.
- 21: PSYC , Site none, Mon 03-Feb-1992 16:19:38 GMT, PSYC CHANGE, 10m with 2m:
PSYC CHANGE, 10m with 2m:
The 10m psyc was changed with
the 2m psyc at 15:40 GMT.
- 22: PSYC , Site none, Mon 03-Feb-1992 17:39:26 GMT, Moved pressure port.
Moved pressure port.
Pressure port was on 4M psyc.
Moved to 2M psyc.Approx 1700 Z.
- 64: PSYC , Site none, Sun 09-Feb-1992 00:27:03 GMT, EGG SIGNALS:
EGG SIGNALS:
At this time the EGG signals are not
being collected. They will be hooked up
tomorrow.
- 69: PSYC , Site none, Mon 10-Feb-1992 14:55:07 GMT, 2m HUMITTER
2m HUMITTER
Julian date 41. The 2m humitter
went south for about 5 hours this morning.
From about 5z to 10z. Then appeared to get
its bearings and fell back into line.
- 71: PSYC , Site none, Mon 10-Feb-1992 23:47:41 GMT, EGG t.
EGG t.
The EGG t is now running.
There is still some noise on both it
and on the dp.Also the t seems to be
a little low compared to the 1m psyc.
The problem was that the filter amplifier
was being driven high by its switch chip.
- 88: PSYC , Site none, Wed 12-Feb-1992 21:01:07 GMT, EGG
EGG
The EG&G dewpointer is now operating.
Covar started with new data started at approx.
20:45z.Wires were switched inaccurately from
the 10mv signal to the 0-5v signal around 16z.
Were rewired correctly at approx 20z.Between
20z and 2045, the preprocessing of the data was
altered in prep.config. The old data for channels
113 and 114 on ragwort are as follows.
s=ragwort:113 c=linear(0.01525925,0.0)
s=ragwort:114 c=linear(0.01525925,0.0)
- 106: PSYC , Site none, Tue 18-Feb-1992 23:56:21 GMT, Removed 10m psychrometer and cleaned hu
Removed 10m psychrometer and cleaned humitter
element in alcohol. Replaced on mast and
it is now working.
- 113: PSYC , Site none, Wed 19-Feb-1992 23:39:11 GMT, Replaced psyc.1am with PAM psyc
Replaced 1m psychrometer (psyc.1am) with PAM intercomparison
unit number 7 psychrometer at jd 050 23:24
- 119: PSYC , Site none, Fri 21-Feb-1992 22:33:09 GMT, pam IC intercompare started
finished pam intercomparison number 7 calibration.
Replaced unit with intercomparison unit number 55
at 2222 GMT
- 128: PSYC , Site none, Mon 24-Feb-1992 18:10:00 GMT, Cleaned and replaced filter on 2m humit
17:45 gmt, jday 55
Brought down 2 meter humitter (psyc.2m).
Cleaned the transducer element with a cotton swab and
alcohol. Cut out a new Gore-Tex cover and replaced
it.
Hung it back up by 18:05
- 131: PSYC , Site none, Mon 24-Feb-1992 23:33:21 GMT, Removing the Asman intercomparison unit
Removing the Asman intercomparison unit number 55
At 1730 cst I remove the Asman intercomparison unit to take it to Kansas City
tomorrow. On Wednesday I will bring back another unit for
intercomparison with the ASTER array.
- 133: PSYC , Site none, Tue 25-Feb-1992 16:12:17 GMT, Attached PAM psychrometer #35 to psyc.1
Attached PAM psychrometer #35 to psyc.1am at 16:08 gmt
There has been no unit attached to psyc.1am since yesterday
morning.
- 137: PSYC , Site none, Wed 26-Feb-1992 18:31:50 GMT, Maintainence of 10m humitter (unit 51)
Took down 10m humitter (unit 51) at 17:50Z, day 057.
Disconnected plugs on boom to let any
moisture evaporate.
Cleaned transducer element with alcohol.
Replaced GoreTex filter wrap.
Replaced unit on tower at 18:13 Z, and it
is now working.
- 142: PSYC , Site none, Thu 27-Feb-1992 22:47:11 GMT, SWITCHING HUNCHROMETERS 1am for 10m
SWITCHING HUNCHROMETERS 1am for 10m
Between 1620-1640 cst (2220-2240z) the humchrometers at station 1am and
10m were interchanged.
- 146: PSYC , Site none, Fri 28-Feb-1992 21:58:15 GMT, Replaced PAM pysc #51 with #55
Removed PAM psychrometer unit #51 at psyc.1am.
Replaced it with unit #55, at 21:50Z, day 59.
#55 has been here before - but got rain damaged.
Be sure to remove it if (it ever) rains!
- 149: PSYC , Site none, Sun 01-Mar-1992 16:29:30 GMT, IC unit 55:
IC unit 55:
The humitter on IC 55 was cleaned with
alcohol and a Vaisala filter added to it.
This occurred at 16:20 GMT.
- 161: PSYC , Site none, Tue 03-Mar-1992 21:10:27 GMT, IC unit 55:
IC unit 55:
The PAM intercomparison unit has been
removed from the array at about 21:05Z.
- 162: PSYC , Site none, Tue 03-Mar-1992 22:15:58 GMT, INC unit 55:
INC unit 55:
The PAM IC unit 55 was hooked backup
on psyc.1am about 10 minutes ago; 22:05Z.
- 174: PSYC , Site none, Fri 06-Mar-1992 14:18:47 GMT, 7M PSYC DOWN & 2M RH ?:
7M PSYC DOWN & 2M RH ?:
The 7m psyc went down last night
at about 3:30Z. Power was recycled and it
came back to life, 14:15Z.
The 2m rh looked funny throughout
the night. It is looking good now.
- 180: PSYC , Site none, Sun 08-Mar-1992 14:45:23 GMT, PSYC IC UNIT:
PSYC IC UNIT:
IC unit 32 was added to network
last night around 00:00Z.
- 183: PSYC , Site none, Sun 08-Mar-1992 17:17:26 GMT, 2M PSYC:
2M PSYC:
Like the 1m prop, the 2m psyc
fixed itself while being examined. There
was no visible dirt on the transducer, and
the transducer was firmly attached.
- 191: PSYC , Site none, Mon 09-Mar-1992 21:11:56 GMT, 2M RH MISBEHAVING:
2M RH MISBEHAVING:
The 2m rh is jumping around.
Looks like a connector problem.
- 192: PSYC , Site none, Mon 09-Mar-1992 22:53:41 GMT, 2M PSYC FIX:
2M PSYC FIX:
The connector between the humitter
transducer and its electronics was intermittent.
When the sheild was tighten down all the way it would
move the transducer. The sensor looks good now.
Installed at 22:45Z.
- 198: PSYC , Site none, Tue 10-Mar-1992 23:07:44 GMT, 2M PSYC STATUS:
2M PSYC STATUS:
The 2m rh started acting up again.
So major surgery was done on the humitter.
The patient died!
We will try to get our spare psyc back from
PAM. The broken psyc will be installed so
we are getting temperature data.
Installed at 23:00Z.
- 202: PSYC , Site none, Wed 11-Mar-1992 20:56:30 GMT, 2M RH RUNNING:
2M RH RUNNING:
A new psychrometer was installed
in the 2m beehive. Sensor id 58 was replaced
with 51. This occurred at about 20:00Z.
- 208: PSYC , Site none, Sat 14-Mar-1992 14:23:42 GMT, PSYC IC UNIT REMOVED:
PSYC IC UNIT REMOVED:
The psyc ic unit was removed
yesterday for PAM IC tests. It will not
be put back on. Removal occurred at about
15:00Z.
- 214: PSYC , Site none, Sun 15-Mar-1992 16:14:09 GMT, PSYC IC TEST:
PSYC IC TEST:
A special psychrometer intercomparison
test was started at 16:10Z. All psycs are at
the 1m height. The test will continue until
tomorrow. Archiving and covars are running.
- 233: PSYC , Site none, Wed 21-Jul-1993 17:48:19 GMT, Don't trust some temperature/humidity p
Don't trust some temperature/humidity profiles.
After trying to get the Bowen ratio and flux-profile relations to
compare reasonably with eddy-correlation flux estimates, I have
found that several of the T/RH profile levels are suspect.
fun.Tddiff.plot(50) shows a case when the 10m temperature is bad and
fun.RHdiff.plot(50,7) shows that the 2m RH is bad during this period, thus
fun.qdiff.plot(50,.5) shows bad q's from 2m and 10m.
Also, it is not too obvious, but fun.RHdiff.plot(60) shows the 4m RH is
slightly off, and thus crosses zero early at 15Z. The Bowen ratio calculated
from (T7-T1)/(q7-q1) agreed much better than (T7-T4)/(q7-q4) with the
eddy-correlation ratio w'T'/w'q' for this period. (Of course, "T" here is
potential temperature.) Thus, I ended up using 1,4,7 for flux-profile
relations and just 1,7 for Bowen ratios...Steve